Ozone Gas Analyzers |

|
OzoMeterTM |
|
Typical Applications
· Fully Featured · Fast and Continuous Measurement · High Accuracy · Direct Reading · Low Maintenance · Keyboard Programmable |
|
Standard Features
¨ Simple keypad control ¨ A bright easy to read LED 5 digit data display, coupled with a LCD function/diagnostic/calibration display ¨ U/V light absorption ¨ Multiple display output units for ozone concentration ¨ Automatic temperature and pressure compensation ¨ 4-20 mA output signal proportional to ozone concentration (Optional second 4-20 mA output available) ¨ Field adjustable alarm setpoint and dry alarm contacts (Optional second alarm setpoint available) ¨ NEMA 12 (standard) enclosure for wall mounting (NEMA 4 or 4X optional)
|

|
Models |
Range |
|
HA-20 MGTP |
0-20 mg/Nm3 (0-20 ppm vol) |
|
HA-100 GTP |
0-95 g/Nm3 (0-8.0% by wt) |
|
HA-200 GTP |
0-190 g/Nm3 (0-16.0% by wt) |
|
Accuracy: |
± 3% of reading |
|
Zero drift: |
0% day |
|
Span drift: |
0% day |
|
Display: |
5 digit + 16 characters |
|
Repeatability: |
1% |
|
Cycle time: |
12 seconds every 2 min. |
|
Rise time: |
8 seconds |
|
Fall time: |
8 seconds |
|
Linearity: |
Better than 1% |
|
Temp. compensation range: |
10-50o C (50o - 120o F) |
|
Pressure compensation range: |
50-200 kPa absolute (0.5-2 ATM) |
|
Sample flow: |
0.5-3 L/min (1-6 cfh) |
|
Power: |
120V, 50/60 Hz, 0.7A |
|
Alarm output: |
4 PDT dry contacts 2A @ 120 VAC |
|
Recorder output: |
(Optional second alarm available) 4-20 mA (Optional second output available) |
|
Dimensions: |
(Optional RS-232C communications port available) 45 cm High x 35.5 cm Wide x 15 cm Deep (16”x14”x6”) |
|
Weight: |
11.4 kg (25 lbs) |
|
Connections: |
1/4” NPT female, 316 SS |